Concept

Instrument — where it appears

A device that reports a colour as numbers — a spectrophotometer, a colorimeter, a densitometer. Each carries its own geometry, bandpass and colour-matching functions, so a reading is a statement about the instrument as much as about the sample, and two instruments agree only where those three do.

Named by 8 essays across one field — each of them below, with the objects they name alongside it.

What an instrument's slit width does to a tabulated colour. The horizontal axis is the full width of a triangular slit, from zero — perfect point sampling — to twenty nanometres; the vertical is the distance from the true colour, logarithmic. For a smooth light the lines are flat: a slit narrower than any feature changes nothing. For a line spectrum they fall off a cliff at the left. Point-sampling a mercury line at five nanometres costs 1.26 ΔE₀₀ and integrating the same spectrum through a five-nanometre slit costs 0.014. A spectrometer does not sample a spectrum; it integrates one, and the blur everybody would remove if they could is what makes a five-nanometre table safe.

The slit is what makes it legal

Point-sampling a mercury line at five nanometres costs a colour difference of one unit, and a laser projector thirty-seven. Integrating the same spectrum through the five-nanometre slit every spectrometer already has costs 0.014 and 0.19. The blur anybody would remove if they could is what makes a coarse table honest.

light · Light
A 12-nanometre notch at 546.1 nm under a fluorescent tube, tabulated five ways. The cost against a tenth-nanometre reference, on a five-nanometre grid, of a notched sample under a fluorescent tube, mercury lines on a phosphor bed, when the two factors of the colour are tabulated as points, when the lamp alone is measured through a five-nanometre slit, when the sample alone is, when each is measured through its own slit, and when the light the sample reflects is measured through one slit. The costs are 2.535 for both sampled at points, 0.379 for the lamp through a slit, 2.724 for the sample through a slit, 0.727 for both through their own slits, 0.019 for the product through one slit. The tick on the two-slit bar is the same two tables summed at a tenth of a nanometre, 0.749: what the separate slits leave is not the grid's.

Two slits are not one slit

A spectrometer's slit is what makes a coarse table honest, for a lamp and for a notched sample alike. But a colour is a sum over the product of the two, and a notch measured through one slit and a lamp measured through another are not the product measured through a slit. Under a smooth light the difference is nothing. Under a fluorescent tube a notch on the mercury line comes out 0.73 colour differences off from two slits — worse than no slit at all for some notches — and 0.02 off from one slit on the reflected light.

light · Light
Six ways of tabulating one notch on one mercury line. A 12-nanometre notch centred on a fluorescent tube's 546.1 nm line, its colour computed on a five-nanometre grid six ways, on a logarithmic scale. Point sampling costs 2.54 colour differences and two separate slits 0.727. Sharpening both blurred tables with the published three-term correction takes it to 0.188 — a real improvement, four times better — and one slit on the light the sample actually reflects gives 0.019. The correction recovers the part of the damage that is a blur, and the part that is left is not a blur.

A linear repair for a bilinear loss

The Stearns correction sharpens a table blurred by a triangular slit, and the obvious question was whether applying it to a lamp's table and a sample's restores the product the two of them are wrong about. It restores four fifths of the damage and cannot touch the rest: a three-term filter is linear, the covariance two separately blurred tables discard is bilinear in the two factors, and no linear operator applied to each factor separately produces a bilinear term. What is left is ten times the one-slit answer, at every position of the notch.

light · Light
One instrument, one slit, two requirements. The slit's width swept, with three measurements on a logarithmic scale: a smooth sample under the line lamp, where only the lamp's structure is at stake; the notched sample under a smooth lamp, where only the notch is; and the real case, both at once. The lamp wants a slit of 5 nanometres and the sample wants 1, and each wants what it wants for the same reason: a slit should spread a feature the grid cannot resolve and leave one it can. The real case is best at 3 nanometres — which is neither requirement's answer — and costs 0.247 there, an order of magnitude more than either requirement alone.

One slit, two requirements

A line lamp wants a wide slit, because a wide slit spreads a line where a coarse grid can see it. A notched sample wants a narrow one, because a wide slit fills the notch the grid could have resolved. An instrument has one slit. Measured on the two requirements separately the best widths are five nanometres and one; measured on the two together the best is three, which is neither — and it costs twelve times what the lamp alone would cost and thirty-four times what the sample alone would.

light · Light
An interference notch filter, and where a five-nanometre grid lands on it. The transmittance of a Fabry-Pérot etalon of order 24 and finesse 20, drawn at a fifth of a nanometre, with the standard grid's points marked. Its features are 2.29 nanometres wide and spaced 22.9 apart, so the grid steps over them: between two adjacent grid points the transmittance rises and falls completely, and neither point records it. That is what a real coating looks like, and a Gaussian notch — which is what this collection's earlier work used — is a much gentler object.

A finer table is a worse table

A real interference filter is not a Gaussian notch. It is an etalon, with pass bands a nanometre or two wide spaced twenty-three apart, and a five-nanometre grid steps over them. Resampled from the maker's one-nanometre table its colour is out by five colour differences under every fill-in rule — the three rules agree to three decimals, because none of them is ever handed a sample inside a feature. The same filter measured through a five-nanometre slit is out by 0.03.

light · Light
Every pair of slits, over 68 notches. The colour error, in ΔE₀₀ from the truth, for every pair of slit widths — the lamp's table blurred through the width down the side, the sample's through the width across — on 68 notches, the mean over all of them under a fluorescent tube. Circle area follows the error. The best pair is 5 nm on the lamp and 1 nm on the sample, at 0.26; the best single slit, on the diagonal, is 5 nm at 0.36. One slit on the reflected light, at 5 nm, averages 0.016.

A second slit buys a quarter

A line lamp wants a five-nanometre slit and a notched sample a one-nanometre slit, so an instrument with a slit for each should do much better than one with a single compromise. Over sixty-eight notches under a fluorescent tube, it does better by 28 per cent. One slit on the reflected light does fifteen times better than the best pair, and an oracle choosing the best pair for every notch is still six times worse. The error was never that the factors were flattened; it was that they were flattened separately.

light · Light
Three bounds against the error they bound, over 68 notches under a fluorescent tube. Each notch placed across by its actual colour error from blurring the lamp and the sample separately, and up by a bound on that error, both on logarithmic scales; the dashed diagonal is where a bound equals the error, and a valid bound sits above it. Cauchy–Schwarz with the true window variances is above the diagonal on every notch, a median 14.7 times the error. Estimated from the blurred tables it falls below on 6 of 68, as low as 0.45 of the error. The Bhatia–Davis bound from the tables and declared ranges is above on every notch and a median 196 times the error.

The tables cannot bound what they discarded

A colour computed from a lamp's blurred table and a sample's blurred table is wrong by the covariance the two blurs threw away, and Cauchy–Schwarz bounds a covariance by two variances. With the true variances the bound always holds and sits fifteen times above the error. With variances read from the tables it fails on six of sixty-eight notches under a fluorescent tube and twenty-two under a laser projector — on the line, where the error is largest. A blurred table does not carry the width of a line, and the covariance depends on it.

light · Light
What declaring a narrowest feature buys, and where it stops being true. The median looseness of a Cauchy–Schwarz bound whose lamp variance is bounded by a declared narrowest feature, against the width declared, for a fluorescent tube and a three-laser projector. Each lamp's own Bhatia–Davis bound — the peak declared and nothing else — is the upper dashed line, and the bound with the true variances is the lower one. The marks are the width each lamp's lines actually have. Declaring it truly takes the tube from ×196 to ×86 and the projector from ×30 to ×14. The open circles are declarations the lamp does not meet, where the bound falls below the error.

A declared width buys a factor of two

A colour engine given two separately blurred spectral tables cannot bound its own error from them, and the bound that always holds — the peak declared and nothing else — sits a median 196 times above the error under a fluorescent tube. Adding one number, the width of the lamp's narrowest feature, brings that to 86. It never fails on any declaration the lamp truly meets, it fails on 47 of 68 notches on one it does not, and its rank correlation with the error it bounds is 0.27.

light · Light

Named alongside it

The objects these essays reach for when they reach for this one.

Spectral structureSpectrophotometryWavelength gridBandpassFluorescentIntegrationMeasurement conditionTransmittanceAliasingSpectral resolutionBilinearityConvention