Spectral structure — where it appears
Named by 24 essays across 5 fields — each of them below, with the objects they name alongside it.
The illuminant is half the answer
An object has a reflectance, not a colour. The colour appears when a specified light falls on it, which is why two surfaces can match in a shop and clash outside, and why every serious matching standard names the light.
Which lamp changes are free
The changes of light that existed before electricity commute with one another to a couple of parts in a thousand, so one set of axes handles all of them. The lights the lighting industry invented do not, and the worst pair in the census is seventy-six times further from commuting than the best.
Three numbers cannot see a line
An instrument that returns three filtered readings of a spectrum determines a three-dimensional projection of it and is exactly blind to the other seventy-eight. On daylight that costs almost nothing; on a fluorescent tube, three quarters of the lamp lies in the part no reading reaches, and adding filters recovers it slowly.
The slit is what makes it legal
Point-sampling a mercury line at five nanometres costs a colour difference of one unit, and a laser projector thirty-seven. Integrating the same spectrum through the five-nanometre slit every spectrometer already has costs 0.014 and 0.19. The blur anybody would remove if they could is what makes a coarse table honest.
A finer reading of a coarser table
Interpolating a five-nanometre spectrum to one nanometre helps a daylight calculation by a factor of five and harms a three-emitter LED by a factor of a hundred and twenty thousand. Both are the same operation on the same table, and which one happens is decided by a property of the light nobody records.
Where the grid starts
Holding the step at five nanometres and sliding the grid's origin through one cell moves a fluorescent tube's computed colour by 3.18 ΔE₀₀ and a laser projector's by 35.0. Refining the step does not fix it and averaging over origins hides it. It is the one tabulation fault with no smooth error to cancel against.
Which end to buy
Refining a five-nanometre grid to one buys a daylight calculation 0.05 ΔE₀₀ and widening its range buys 0.54. Under a fluorescent tube the same two purchases are worth 0.83 and 0.0001. The ranking reverses completely, and what decides it is one length compared against one other length.
The lens is worst under tungsten
An ageing lens costs 4.38 ΔE₀₀ under a tungsten lamp and 2.20 under a fluorescent tube. The pigment peaks cost 2.84 under a three-emitter LED and 1.87 under the same tube. Six departures across six lights do not form a single ordering, which is what a pairing looks like and what a dominant term would not.
The macular is a band, not a filter
The lens absorbs everything below 500 nanometres with a long tail; the macular pigment absorbs forty nanometres either side of 460 and nothing else. The two have similar sizes and completely different distributions, and the reason is that one is broad and one is narrow — which is what decides whether a sample is affected at all.
A brand colour for a population
A brand colour is chosen once, specified in a number, and seen by everybody. The observer audit says two candidate colours with identical specifications can carry observer spreads differing by an order of magnitude, and that the spread is computable from an ink's spectrum at the moment the ink is chosen.
A lamp is two audits at once
A lamp's ultraviolet content decides what its tabulation costs and its blue content decides what its observer costs, and the two run in opposite directions with colour temperature. So no lamp is good for both audits and no lamp is bad for both, and a single figure of merit for either is a figure of merit for one property of a spectrum.
The line can be in the sample
The rule for which tabulation defect to fix compares the light's narrowest feature with the grid's step, and it named its own failure case — a sample with structure narrower than the step. Given one, a two-nanometre notch under a thermal source with no feature at all costs 2.1 colour differences at five nanometres and moves 2.0 when the grid slides, which is what a fluorescent tube costs on a smooth pigment. Under that tube a notch twelve nanometres wide, more than twice the step, moves 8.1 when it sits on the mercury line.
The cost of a steep notch repeats every step
A Gaussian notch is safe on a five-nanometre grid once it is a couple of steps wide. A flat-bottomed notch with steep sides never is. Its cost on the grid rises and falls with its width, with the step as its period — 0.02 colour differences at ten nanometres, 1.99 at twelve and a half, 0.03 at twenty — and it does not die away as the notch widens. What sets its size is how fast the edges rise, and an interference filter's edges rise in under a nanometre.
Two slits are not one slit
A spectrometer's slit is what makes a coarse table honest, for a lamp and for a notched sample alike. But a colour is a sum over the product of the two, and a notch measured through one slit and a lamp measured through another are not the product measured through a slit. Under a smooth light the difference is nothing. Under a fluorescent tube a notch on the mercury line comes out 0.73 colour differences off from two slits — worse than no slit at all for some notches — and 0.02 off from one slit on the reflected light.
The room a surface needs is written in its band
Whether two yellow filters cancel on a surface depends on the room, and each surface has its own crossing — the degree of adaptation at which the shared yellowing falls to the size of what is left underneath. Those crossings run from 0.68 to 0.99, and where a surface's absorption band sits accounts for almost all of the spread while how much light it returns accounts for almost none. The reds need a room brighter than a graphic-arts viewing booth, which is brighter than any room a sample is judged in.
Flicker sorts lamps the wrong way
A camera cannot see a spectral line in its own white, so the essay on the two-matrix profile named the classifiers a device might have instead, and the first of them was flicker. Flicker is measurable, and it measures the wrong thing. It sorts lamps by how their power is delivered while the matrix needs them sorted by how their spectrum is shaped, and the two are independent: a white LED on a constant driver is perfectly steady and strongly structured, and it is what most indoor photographs are lit by.
A linear repair for a bilinear loss
The Stearns correction sharpens a table blurred by a triangular slit, and the obvious question was whether applying it to a lamp's table and a sample's restores the product the two of them are wrong about. It restores four fifths of the damage and cannot touch the rest: a three-term filter is linear, the covariance two separately blurred tables discard is bilinear in the two factors, and no linear operator applied to each factor separately produces a bilinear term. What is left is ten times the one-slit answer, at every position of the notch.
One slit, two requirements
A line lamp wants a wide slit, because a wide slit spreads a line where a coarse grid can see it. A notched sample wants a narrow one, because a wide slit fills the notch the grid could have resolved. An instrument has one slit. Measured on the two requirements separately the best widths are five nanometres and one; measured on the two together the best is three, which is neither — and it costs twelve times what the lamp alone would cost and thirty-four times what the sample alone would.
A finer table is a worse table
A real interference filter is not a Gaussian notch. It is an etalon, with pass bands a nanometre or two wide spaced twenty-three apart, and a five-nanometre grid steps over them. Resampled from the maker's one-nanometre table its colour is out by five colour differences under every fill-in rule — the three rules agree to three decimals, because none of them is ever handed a sample inside a feature. The same filter measured through a five-nanometre slit is out by 0.03.
Two sensors disagree about deep red, not lines
A phone's ambient-light sensor and its camera read the same lamp differently, and the difference sorts fourteen lamps into smooth and structured without a single mistake — where flicker made five. It is not reading their lines. Almost all of it comes from the sensor's clear channel collecting deep red the camera's infrared cut throws away, so daylight with its far red trimmed is called structured and a white LED with a far-red emitter is called smooth.
A second slit buys a quarter
A line lamp wants a five-nanometre slit and a notched sample a one-nanometre slit, so an instrument with a slit for each should do much better than one with a single compromise. Over sixty-eight notches under a fluorescent tube, it does better by 28 per cent. One slit on the reflected light does fifteen times better than the best pair, and an oracle choosing the best pair for every notch is still six times worse. The error was never that the factors were flattened; it was that they were flattened separately.
The tables cannot bound what they discarded
A colour computed from a lamp's blurred table and a sample's blurred table is wrong by the covariance the two blurs threw away, and Cauchy–Schwarz bounds a covariance by two variances. With the true variances the bound always holds and sits fifteen times above the error. With variances read from the tables it fails on six of sixty-eight notches under a fluorescent tube and twenty-two under a laser projector — on the line, where the error is largest. A blurred table does not carry the width of a line, and the covariance depends on it.
A declared width buys a factor of two
A colour engine given two separately blurred spectral tables cannot bound its own error from them, and the bound that always holds — the peak declared and nothing else — sits a median 196 times above the error under a fluorescent tube. Adding one number, the width of the lamp's narrowest feature, brings that to 86. It never fails on any declaration the lamp truly meets, it fails on 47 of 68 notches on one it does not, and its rank correlation with the error it bounds is 0.27.
A narrow channel has to be read on its own
An ambient sensor with a clear channel tells structured lamps from smooth ones by reading deep red, and one without it reads structure but breaks at one per cent of calibration error. A narrow fourth channel between the camera's peaks was proposed to have both. Pooled into the sensor's disagreement with the camera, it reads structure and breaks at one and a quarter per cent — no better than the design it was meant to rescue. Read on its own, as its departure from what the camera predicts for it, a channel at 450 nm holds to four per cent, the clear channel's figure, and far red does not move it.
Named alongside it
The objects these essays reach for when they reach for this one.
Wavelength gridFluorescentSpectrophotometryBandpassInstrumentAliasingMeasurement errorSpecificationIntegrationObserver metamerismQuadratureTransmittance